Function and Use
The moisture meter is a new type of rapid moisture detection instrument. Compared with the international oven heating method, its test results have good consistency, are interchangeable, and the detection efficiency is much higher than the oven method. It only takes a few minutes to complete the measurement of general samples. The instrument is easy to operate and the test is accurate. The display part uses a red digital tube, and the display value is clear and visible. It can display the moisture value, sample initial value, final value, measurement time, temperature initial value, final value and other data. And it has the function of connecting to a computer and a printer.
Compliant with standards
GB/T26497-2011, QB/T1671
Test principle
The ring-shaped halogen heater ensures that the sample is evenly heated during the high-temperature test, making the sample surface less susceptible to damage and drying quickly. During the drying process, the moisture meter continuously measures and instantly displays the sample's lost moisture content%. After the drying process is completed, the final measured moisture content value is locked and displayed.
Instrument features
1. All-metal anti-oxidation cast aluminum shell design, extremely high corrosion resistance
2. RS-232 bidirectional data transmission and USB dual interface facilitate users in various ways
3. The result printing of connecting peripheral devices complies with GLP specifications, and the time and date can be set manually, which is convenient for users to better control quality
4. There is a Kingston safety lock hole at the back of the product
5. Three heating modes are available, suitable for different samples in various industries
6. Up to 49 detection modes and 99 sets of test data can be stored
7. Ultra-large backlit LCD display design, range tracking indicator bar, convenient for users to intuitively view the range usage
8. Four languages are available: English, French, Spanish and German
9. Humanized scale bracket design makes it easier for users to remove samples
Technical parameters
Range: 50g
Readability: 0.001g / 0.01%
Repeatability error: 0.002g /0.05% (10g sample)
Stabilization time: 4 seconds
Scale size: Ø 100mm
Heating method: 400W ring halogen lamp for uniform heating
Temperature range: 50℃ -160℃
Heating mode: standard heating/step heating/ramp heating
Power supply: 110V or 220V, 50-60Hz.
Detection mode: up to 49 detection modes can be stored
Test results: up to 99 sets of data can be stored
Operating temperature: 0℃-40℃
Overall size: 250mm×360mm×185mm
Input/output: RS-232 bidirectional data transmission interface USB interface
Result display method: moisture percentage, solid percentage, moisture-solid ratio, solid-moisture ratio
Stop method: manual stop, automatic stop after specified time, automatic stop after stable reading
Instrument configuration
Host, other accessories