Function and Use
This machine is used to detect trace moisture in solvents. It has high detection accuracy, simple operation, and direct digital display of results. It is mainly used to determine the moisture content in solvents, paper, oil products, medicines, coatings and other products. The instrument is advanced and easy to operate, and the test data is reliable and accurate. It is suitable for the determination of moisture content in many industries such as pharmaceuticals, fine chemicals, petrochemicals, scientific research, colleges and universities, food, tobacco, light industry, etc.
Scope of application: organic compounds generally saturated, unsaturated hydrocarbons, acetals, acyl halides, acids, alcohols, stable acyl, amides, weak amines, anhydrides, disulfides, lipids, ether halides, peroxides, orthoesters, salts and thioethers.
Inorganic compounds generally, acidic oxides, aluminum oxide, acyl, barium peroxide, calcium carbide, copper oxide, desiccant, hydrazine sulfate, and salts of some organic and inorganic acids.
Comply with standards
GB/T 7600-1987, GB 6283-1982, SH/T 0246, GB/T 11133-1989, GB/T 1600, GB/T 11146,
GB/T 12717, GB/T 5074
Test principle
The moisture analyzer uses Karl Fischer volumetric titration to determine the moisture content in the sample being tested.
Instrument features
1. Full microcomputer control, automatic measurement; fast measurement speed, high accuracy, good repeatability
2. Automatic liquid aspiration, automatic liquid injection, automatic measurement, automatic waste liquid discharge
3. Data storage function, measurement results can be transferred to the computer; multi-speed stirring system;
4. LED display, friendly display interface, real-time display of measurement process, easy operation
Technical parameters
Measurement method: volumetric titration
Measurement range: 0.001%-100%
Resolution: 0.05ml
Titration flow: automatic control
Indication error: <1%
Power supply voltage: AC 220V